Posted in | News | Laser | Medical Optics

Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays

In an article recently published in Scientific Reports, researchers presented a novel approach to enhance the sensitivity of lateral flow assays (LFAs) by utilizing laser micromachined constraints in nitrocellulose (NC) membranes. They addressed the critical need for improved diagnostic tools, particularly during viral outbreaks, by proposing a method that significantly boosts LFA performance.

Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays

Image Credit: steved_np3/Shutterstock.com

The study focused on increasing the colorimetric sensitivity of LFAs, which are widely used for point-of-care (PoC) diagnostics, especially during outbreaks like severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).

Introduction to Lateral Flow Assays

LFAs are widely used diagnostic tools known for their simplicity, speed, and affordability. They are commonly employed in PoC settings to detect various pathogens, including viruses like SARS-CoV-2, bacteria, and parasites, as well as other medical conditions.

Despite their advantages, traditional LFAs have limitations in sensitivity and quantitative analysis. They often require high analyte concentrations for accurate detection. Enhancing the sensitivity of LFAs is crucial for early and precise diagnosis, mainly during viral outbreaks.

Technological Advancements in LFAs

LFAs typically consist of several components, including a sample pad, conjugate pad, NC membrane, and absorbent pad. The NC membrane is important in facilitating capillary flow and enabling specific reactions between analytes and antibodies.

Recent advancements have aimed at improving the limit of detection (LoD) and sensitivity of LFAs through various strategies, such as sample pretreatment, flow rate modification, and new colorimetric methods. Laser micromachining has emerged as a promising technique to enhance LFA performance by creating microchannels in the NC membrane, which control flow rate and reaction kinetics.

Enhancing Sensitivity through Laser Micromachining

This paper aimed to improve the sensitivity of LFAs by modifying the flow dynamics within the NC membrane. The authors employed femtosecond laser micromachining to create microchannels in the NC membrane, intending to prolong the interaction time between the analyte and the labeled antibody. This approach was hypothesized to increase specific binding interactions and, consequently, the signal intensity of the assay.

The experimental setup utilized the second harmonic of Ytterbium-doped Potassium Gadolinium Tungstate (Yb: KGW) femtosecond laser to ablate microchannels of varying widths and densities in the NC membrane. The effect of these microchannels on vertical wicking speed was evaluated through video recordings.

Additionally, the researchers synthesized gold nanoparticles (AuNPs) to mimic LFA operation and their impact on wicking flow and SARS-CoV-2 detection sensitivity.

Key Findings and Insights

Laser micromachining of NC membranes significantly enhanced the sensitivity of LFAs. The creation of microchannels in the NC membrane resulted in a controlled delay of analyte flow, increasing interaction time by up to 950 %. This interaction led to a 40 % improvement in colorimetric signal intensity compared to standard NC LFAs.

The authors observed that the width and length of the microchannels played crucial roles in modulating flow dynamics. Wider and longer microchannels consistently reduced wicking velocity, thereby increasing reaction time and improving assay sensitivity. For instance, microchannels with a width of 50 μm and a length of 20 mm exhibited the most significant improvement in signal intensity.

The study also highlighted the importance of optimizing laser ablation parameters to achieve precise microchannel structures without damaging the NC membrane. "Slow" ablation conditions, characterized by lower scanning speeds and higher pulse energies, were more effective in creating channels that minimized flow turbulence and resistance.

Applications

The enhanced sensitivity of LFAs achieved through laser micromachining has significant implications for PoC diagnostics. The improved signal intensity and reaction kinetics can lead to more accurate and reliable detection of low-concentration analytes, making LFAs more effective for early-stage disease diagnosis. This advancement is significant for detecting respiratory viruses like SARS-CoV-2, where rapid and sensitive diagnostic tools are crucial for controlling outbreaks.

The laser micromachining technique offers a cost-effective and scalable solution for manufacturing high-sensitivity LFAs. The ability to precisely control flow dynamics within the NC membrane opens new possibilities for developing advanced diagnostic assays with multiplexing capabilities, enabling the simultaneous detection of multiple pathogens from a single sample.

Conclusion and Future Directions

The study concluded that femtosecond laser micromachining is a promising technique for enhancing the sensitivity of LFAs. By creating precise microchannels in NC membranes, researchers significantly improved the immunological reaction time and signal sensitivity of the assays. This approach can potentially revolutionize diagnostics, particularly for viral infections like SARS-CoV-2, by providing accurate and reliable results.

Future work could explore integrating this method with other signal enhancement strategies, such as using fluorescent nanoparticles or advanced optical detection methods, to further improve LFA performance. Additionally, the scalability and practical implementation of laser micromachining in commercial LFA production require further investigation to realize its full potential in clinical diagnostics.

Journal Reference

Khatmi, G., et al. Lateral flow assay sensitivity and signal enhancement via laser µ-machined constrains in nitrocellulose membrane. Sci Rep. (2024). DOI: 10.1038/s41598-024-74407-3, https://www.nature.com/articles/s41598-024-74407-3

Disclaimer: The views expressed here are those of the author expressed in their private capacity and do not necessarily represent the views of AZoM.com Limited T/A AZoNetwork the owner and operator of this website. This disclaimer forms part of the Terms and conditions of use of this website.

Muhammad Osama

Written by

Muhammad Osama

Muhammad Osama is a full-time data analytics consultant and freelance technical writer based in Delhi, India. He specializes in transforming complex technical concepts into accessible content. He has a Bachelor of Technology in Mechanical Engineering with specialization in AI & Robotics from Galgotias University, India, and he has extensive experience in technical content writing, data science and analytics, and artificial intelligence.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Osama, Muhammad. (2024, October 14). Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays. AZoOptics. Retrieved on November 22, 2024 from https://www.azooptics.com/News.aspx?newsID=29996.

  • MLA

    Osama, Muhammad. "Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays". AZoOptics. 22 November 2024. <https://www.azooptics.com/News.aspx?newsID=29996>.

  • Chicago

    Osama, Muhammad. "Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays". AZoOptics. https://www.azooptics.com/News.aspx?newsID=29996. (accessed November 22, 2024).

  • Harvard

    Osama, Muhammad. 2024. Revolutionizing Diagnostics: Laser-Enhanced Lateral Flow Assays. AZoOptics, viewed 22 November 2024, https://www.azooptics.com/News.aspx?newsID=29996.

Tell Us What You Think

Do you have a review, update or anything you would like to add to this news story?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.