Utilizing X-Ray Fluorescence for Glass Analysis

Single component fused silica (SiO2) is the simplest form of glass. However, it must be noted that this material is both highly expensive and very difficult to process. Thus, in an attempt to help minimize these difficulties, some other oxides are added, offering specific properties to the resultant glass.

Most glasses are made up of approximately 70% silica, which is a glass former, lime as a stabilizer in the form of limestone (about 10%), and soda as a flux in the form of carbonate and sulfate (about 14%).

Other types of oxides enhance the physical characteristics of glass, like alumina or magnesia, and particularly improve the resistance to atmospheric conditions. The incorporation of different metallic oxides, such as chromium, iron, manganese or copper oxides, ensures in-depth coloring.

Instrumentation

An ARL OPTIM’X XRF spectrometer from Thermo Electron (Figure 1) has been used within this study to derive limits of precision and detection for the analysis of glass.

As a wavelength dispersive system which offers superior resolution and light elements capability, the ARL OPTIM’X has a maximum power of 50 W and is fitted with an air-cooled Rh End-Window Tube featuring a thin Be window (0.075 mm).

The performance of the ARL OPTIM’X is similar to a 200 W conventional WD-XRF instrument as a result of the close coupling between the X-Ray tube anode and the sample. The instrument can be integrated with the unique SmartGonio, a series of Multichromators, or both.

ARL OPTIM’X XRF Spectrometer.

Figure 1. ARL OPTIM’X XRF Spectrometer. Image Credit: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Limits of Detection & Calibration

An ARL OPTIM’X was used to measure a series of pressed glass samples. Relating intensities for each oxide (or element) to concentrations in standard samples was performed to derive calibration curves.

Though X-Ray fluorescence measures pure elements, the results can be directly related to the oxide forms of such elements (when there is only one single form present in the sample). Limits of detection have been derived utilizing the calibration curves and deploying SmartGonio for the most common oxides/elements found in soda-lime glasses (Figure 2).

Limits of detection for various elements in soda-lime glasses prepared as pressed powders are shown in Figure 1.

Soda-lime glasses.

Figure 2. Soda-lime glasses. Image Credit: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Table 1. Analytical parameters and limits of detection for various oxides/element in soda-lime glass (100 sec. counting time). Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

OXIDE/ELEMENT LINE CRYSTAL DETECTOR LOD [PPM]
Na2O 1,2 AX-06 FPC 100
MgO 1,2 AX-06 FPC 60
Al2O3 1,2 PET FPC 47
SiO2 1,2 PET FPC N.R.
P2O5 1,2 PET FPC 48
SO3 1,2 PET FPC 23
Cl 1,2 PET FPC 24
K2O 1,2 LiF 200 FPC 14
CaO 1,2 LiF 200 FPC 12
TiO2 1,2 LiF 200 FPC 12
Cr2O3 1,2 LiF 200 FPC 9
MnO 1,2 LiF 200 FPC 9
Fe2O3 1,2 LiF 200 FPC 9
ZnO 1,2 LiF 200 SC 3.6
SrO 1,2 LiF 200 SC 2.4
ZrO2 1,2 LiF 200 SC 1.8
BaO 1 LiF 200 FPC 51
PbO 1 LiF 200 SC 9

N.R. = LoD is not relevant for major elements.
FPC = flow proportional counter.
SC = scintillation counter.
Excitation conditions: 40 kV/1.25 mA.
Collimator: 0.29°.

Precision Tests

Repeatedly analyzing the same pressed pellet sample for eleven consecutive analyses was the method used for precision tests. Utilizing a counting time of 36 seconds per analytical line, eighteen elements/oxides were determined.

The results are summarized below for two different glass samples shown in Tables 2 and 3. Counting time can be increased for higher precision. Doubling the counting time results in an improvement in the precision by a factor of about 1.4 (square root of 2).

Table 2. Repeatability for the analysis of the major and minor oxides in sample A. Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

RUN Na2O % MgO % Al2O3 % SiO2 % K2O % CaO % Fe2O3 % SO3 PPM TiO2 PPM P2O5 PPM Cl PPM Cr2O3 PPM MnO PPM As2O3 PPM SrO PPM ZrO2 PPM BaO PPM PbO PPM
Time (s) 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36
Run 1 13.98 0.185 1.79 72.59 0.588 10.85 0.330 582 579 166.9 113.4 93.6 48.4 101.6 127.4 209.1 454.3 228.2
Run 2 13.93 0.193 1.81 72.60 0.582 10.82 0.333 640 563 146.3 129.5 91.4 44.7 101.5 124.8 204.6 392.6 218.9
Run 3 13.97 0.177 1.80 72.64 0.588 10.82 0.330 608 563 193.3 111.2 91.1 43.8 95.9 127.1 207.0 361.8 197.8
Run 4 14.01 0.178 1.80 72.64 0.582 10.87 0.330 645 581 199.2 104.6 96.2 29.9 103.8 127.0 205.4 375.7 234.6
Run 5 13.95 0.182 1.80 72.60 0.588 10.83 0.329 576 564 158.1 111.8 94.6 41.8 103.7 122.7 204.7 385.2 228.5
Run 6 13.94 0.177 1.81 72.61 0.589 10.82 0.329 573 569 171.3 107.9 85.2 49.5 95.0 126.3 203.8 355.4 194.8
Run 7 13.86 0.185 1.80 72.64 0.588 10.83 0.330 658 569 203.6 113.4 88.9 40.3 96.0 125.0 205.4 434.1 234.4
Run 8 13.92 0.186 1.81 72.59 0.585 10.84 0.331 652 566 190.4 135.6 94.5 44.7 96.8 125.9 203.4 315.1 207.3
Run 9 13.94 0.184 1.81 72.63 0.591 10.82 0.334 651 579 150.7 110.1 88.6 43.1 114.9 127.1 206.6 401.2 220.8
Run 10 13.98 0.183 1.80 72.63 0.586 10.87 0.332 617 526 255.0 104.0 83.6 41.9 99.2 125.8 206.2 402.2 214.6
Run 11 13.95 0.188 1.78 72.62 0.588 10.83 0.330 619 561 218.3 97.9 80.6 38.9 97.7 126.8 203.1 429.8 197.8
Avg. 13.95 0.183 1.80 72.62 0.587 10.84 0.331 620 565 186.6 112.7 89.9 42.4 100.5 126.0 205.4 391.6 216.1
Std. Dev. 0.04 0.005 0.01 0.02 0.003 0.02 0.0015 32 15 32.5 11 5 5.2 5.7 1.4 1.8 39.7 15

 

Table 3. Repeatability for the analysis of the major and minor oxides in sample B. Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

RUN Na2O % MgO % Al2O3 % SiO2 % K2O % CaO % Fe2O3 % SO3 PPM TiO2 PPM P2O5 PPM Cl PPM Cr2O3 PPM MnO PPM As2O3 PPM SrO PPM ZrO2 PPM BaO PPM PbO PPM
Time (s) 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36 36
Run 1 13.35 0.180 1.67 73.07 0.556 10.77 773.3 0.177 556.0 200.6 100.7 63.4 9.3 118.2 122.4 227.2 883.9 895.6
Run 2 13.33 0.180 1.68 73.08 0.564 10.76 757.9 0.181 568.0 159.5 111.2 65.9 3.9 112.0 122.2 226.1 960.6 914.5
Run 3 13.28 0.186 1.67 73.08 0.554 10.81 789.6 0.180 555.1 193.3 115.7 64.7 18.2 115.6 119.3 225.3 925.4 911.5
Run 4 13.28 0.185 1.66 73.11 0.559 10.83 768.2 0.186 587.2 156.6 103.5 74.7 8.7 105.6 126.5 225.1 891.3 900.0
Run 5 13.35 0.181 1.67 73.05 0.554 10.79 763.9 0.181 594.7 187.4 97.9 63.6 12.3 106.8 126.3 226.2 948.8 904.8
Run 6 13.32 0.172 1.67 73.11 0.566 10.80 767.3 0.186 541.4 183.0 114.6 59.5 14.7 115.2 125.5 226.3 960.5 904.6
Run 7 13.33 0.185 1.67 73.06 0.554 10.79 758.9 0.180 570.3 193.3 101.8 67.9 6.1 113.5 124.9 226.9 960.6 910.1
Run 8 13.26 0.185 1.69 73.04 0.555 10.78 771.7 0.185 565.2 191.8 95.2 66.9 9.3 101.6 124.7 227.4 918.0 919.7
Run 9 13.33 0.180 1.64 73.11 0.561 10.82 775.7 0.183 553.7 219.7 113.4 61.5 9.3 109.6 124.6 222.9 916.9 913.4
Run 10 13.30 0.193 1.68 73.08 0.556 10.80 764.1 0.188 572.6 243.2 101.3 64.4 1.0 103.4 123.0 226.0 980.8 912.7
Run 11 13.31 0.184 1.66 73.06 0.561 10.78 785.3 0.186 566.0 191.8 109.0 70.8 13.8 106.9 122.1 227.7 950.9 875.6
Avg. 13.31 0.183 1.67 73.08 0.558 10.80 770.5 0.183 566.4 192.8 105.8 65.8 9.7 109.9 123.8 226.1 936.2 905.7
Std. Dev. 0.03 0.01 0.01 0.03 0.004 0.02 10 0.003 15 24.2 7.2 4.3 4.9 5.4 2.2 1.4 31.2 12

 

Conclusion

It was shown for all limits of detection obtained that the ARL OPTIM’X can offer adequate analysis results, significantly for bottle glass application.

For major and minor elements - even for Na2O and MgO - the repeatability of the analysis is high. Should elements present below 100 ppm need to be controlled precisely, a longer counting time may be utilized.

As is shown by these results, the ARL OPTIM’X spectrometer is ideally suited for the generation of precision results regarding the coloring agents in glasses and the determination of the main oxides.

This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers.

For more information on this source, please visit Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers. (2024, February 08). Utilizing X-Ray Fluorescence for Glass Analysis. AZoOptics. Retrieved on November 24, 2024 from https://www.azooptics.com/Article.aspx?ArticleID=2286.

  • MLA

    Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers. "Utilizing X-Ray Fluorescence for Glass Analysis". AZoOptics. 24 November 2024. <https://www.azooptics.com/Article.aspx?ArticleID=2286>.

  • Chicago

    Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers. "Utilizing X-Ray Fluorescence for Glass Analysis". AZoOptics. https://www.azooptics.com/Article.aspx?ArticleID=2286. (accessed November 24, 2024).

  • Harvard

    Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers. 2024. Utilizing X-Ray Fluorescence for Glass Analysis. AZoOptics, viewed 24 November 2024, https://www.azooptics.com/Article.aspx?ArticleID=2286.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.