The Glacier® T series from B&W Tek is a high-resolution, double-pass transmission-based thermoelectric-cooled (TEC) linear CCD array spectrometer particularly designed for Raman spectroscopy.
The Glacier T spectrometer has been preconfigured for 785 nm or 532 nm excitation with a broad spectral range or high-resolution option.
The TE-cooled spectrometer also features a high-speed USB 2.0 interface, a built-in 16-bit digitizer and 2048 elements, thereby constantly delivering improved high-throughput outcomes.
Integrating a Glacier T spectrometer to its equivalent excitation laser system and Raman probe offers an OEM building block or module-level Raman system. In addition, application support and system development are available for OEM applications
Features
- A 16-bit digitizer
- A resolution* of 3.0 cm−1 to 4.5 cm−1
- Rapid F/2 spectrograph
- OEM building block
- Raman shift* of 0 cm−1 up to 4000 cm−1
- 14 °C TE-cooled detector
*-Varies by model
Accessories
- A 532-nm laser
- A 785-nm laser
- An industrial-grade probe
- A lab-grade probe
Raman Spectrum: Cyclohexane.
Image Credit: B&W Tek
Dark Current: Uncooled versus Cooled CCD Detectors at 30 Seconds
Room Temperature. Image Credit: B&W Tek
Cooled to 14 °C. Image Credit: B&W Tek
Specifications
Table 1. Source: B&W Tek
. |
. |
DC Power Input |
5 V DC < 1.5 Amps |
AC Adapter Input |
100 - 240 VAC 50/60 Hz, 0.5 A @ 120 VAC |
Detector Type |
Response Enhanced 2048 element linear silicon CCD array |
Pixels |
2048 x 1 elements @ 14 µm x 200 µm per element |
Spectrograph F# |
2.0 |
Slit |
20 mm (785 nm), 10 mm (532 nm) |
Spectrograph Optical Layout |
Transmission |
Dynamic Range |
300 typical |
Digitizer Resolution |
16-bit or 65,535:1 |
Readout Speed |
500 kHz |
Data Transfer Speed |
Up to 180 spectra per second via USB 2.0 |
Integration Time |
5 ~ 65,535 ms x multiplier |
External Trigger |
Aux Port |
Operating Temperature |
15 °C - 35 °C |
Operational Relative Humidity |
85% Noncondensing |
TE Cooling |
14 °C |
Weight |
~1.8 kg (3.9 lbs) |
Dimensions |
191 mm x 94 mm x 90 mm (7.5 in x 3.7 in x 3.5 in) |
Computer Interface |
USB 2.0 / 1.1 |
Operating Systems |
Windows: XP, 7, 8, 10 (32-bit & 64-bit) |
Available Models
Table 2. Source: B&W Tek
Model # |
Spectral
Range |
Spectral
Resolution |
Approximate
Raman Shift |
Resolution* |
BTC162E-532S |
532 - 676 nm |
~0.15 nm |
0 cm-1 - 4000 cm-1 |
~4.0 cm-1 @ 614 nm |
BTC162E-532H |
532 - 645 nm |
~0.11 nm |
0 cm-1 - 3300 cm-1 |
~3.0 cm-1 @ 614 nm |
BTC162E-785S |
780 - 1050 nm |
~0.37 nm |
0 cm-1 - 3200 cm-1 |
~4.5 cm-1 @ 912 nm |
BTC162E-785H |
784 - 996 nm |
~0.29 nm |
0 cm-1 - 2700 cm-1 |
~3.5 cm-1 @ 912 nm |
*-Varies by Model
Accessories
Raman Probes
Table 3. Source: B&W Tek
Model # |
Grade |
For Excitation |
BAC100-532 |
Lab |
532 nm |
BAC100-785 |
Lab |
785 nm |
BAC102-532 |
Lab (Trigger) |
532 nm |
BAC102-785 |
Lab (Trigger) |
785 nm |
BAC101-532 |
Industrial |
532 nm |
BAC101-785 |
Industrial |
785 nm |
Sampling Accessories
Table 4. Source: B&W Tek
Model # |
Description |
BAC151A |
Video Microscope Sampling System |
BAC150 |
Raman Probe Holder |
BCR100A |
Cuvette Holder |
BAC160 |
Liquid Sample Flow Cells |
Lasers
Table 5. Source: B&W Tek
Model # |
Wavelength |
Power |
BWN-532-OEM |
532 +/- 1 nm |
Up to 100 mW |
BRM-785 |
785 +/- 0.5 nm |
Up to 1 W |